Our website features a wide range of standard power conversion and backup products, but we also have the capacity and capability to design and manufacture custom solutions to meet your specific application needs. Contact our team of experienced engineers today to discuss your project. Dismiss
The MHP650 range meets the IEC60601-1 safety specification for medical equipment.
Three mechanical mounting formats are available to provide design-in flexibility. Two fan-cooled formats provide the option of having the fan mounted internally on the top or externally mounted on the end of the unit. The integral fan speed is load dependent to aid minimizing noise. By using the U channel format designers can supply their own airflow; only 5.5 m per second forced airflow is required. A 12 VDC, 6 W fan supply is provided when using this format.
All units operate from a universal 85-264 VAC input range with full-power available from 90 VAC. The operating temperature range is -20ºC to 70ºC with no derating below 50°C.
Model | Input | Output | Power |
MHP1000PS12 | 80 to 264 Vac | 12 V / 83 A | 1000W |
MHP1000PS15 | 80 to 264 Vac | 15 V / 67 A | 1000W |
MHP1000PS24 | 80 to 264 Vac | 24 V / 50 A | 1000W |
MHP1000PS28 | 80 to 264 Vac | 28 V / 43 A | 1000W |
MHP1000PS36 | 80 to 264 Vac | 36 V / 34 A | 1000W |
MHP1000PS48 | 80 to 264 Vac | 48 V / 25 A | 1000W |
MHP650PS12-EF | 80 to 264 Vac | 12 V / 50 A | 650W |
MHP650PS15-EF | 80 to 264 Vac | 15 V / 40 A | 650W |
MHP650PS24-EF | 80 to 264 Vac | 24 V / 27 A | 650W |
MHP650PS28-EF | 80 to 264 Vac | 28 V / 23 A | 650W |
MHP650PS36-EF | 80 to 264 Vac | 36 V / 18 A | 650W |
MHP650PS48-EF | 80 to 264 Vac | 48 V / 13.5 A | 650W |
Dimensions: | MHP650-EF: 9.18 x 4.00 x 2.50 in (233.2 x 101.6 x 63.5 mm) |
MHP650-TF: 8.00 x 4.00 x 2.58 in (203.2 x 101.6 x 65.5 mm) | |
MHP1000: 9.55 x 5.90 x 2.40 in (242.6 x 149.8 x 61.0 mm) | |
Input Voltage: | 650 W: 80-264 VAC, derate output power 10% <90 VAC, and 20% <85 VAC 1000 W: 85-264 VAC, derate output power 10% <90 VAC |
Harmonic Currents: | EN61000-3-2 class A, 650 W: class C for loads ?20% 1000 W: class C for loads ?10% |
Efficiency: | 85% typical |
Operating Temp: | -20 °C to +70 °C, derate linearly from +50 °C at 2.5% / °C to 50% at +70 °C, 1000 W: -40 °C start up |
Isolation: | 4000 VAC Input to Output 2 x MOPP 1500 VAC Input to Ground 1 x MOPP 500 VDC Output to Ground |
Worldwide Safety Approvals: | IEC60601-1 |
European Safety Approvals: | EN60601-1 Including Risk Management |
US Safety Approvals: | CSA 22.2 No.60601-1, ANSI/AAMI ES60601-1 |
EMC: | EN55022/ENN55011 class B conducted EN55022/EN55011 class A radiated |